TGA Thermogravimetric Analyzer with automatic covers for volatiles in coal, coke and graphite
Multiple sample (19 or 38) automatic analyzer for moisture, volatiles, ashes, fixed carbon and loss on ignition with unique auto cover installation and removal.
Applications
Coal, coke, graphite, iron ore, cement, bauxite ,fertilizers, rubbers, plastics, food, animal feed, snacks, dough, wheat, flour, used engine oil etc...
Complies with AOAC, AACC, ASTM & FDA 21CFR Part 11 Methodologies
For applications with volatiles and covers, it places and removes covers on crucibles automatically inside the furnace: no operator attendance needed, (Patented design), no cool down delay, safe operation.

TGA-2000: Dual Carousel configuration |
|
Technical Data |
|
- Sample size ~1 to 10 g
- Electrical furnace with temperatures up to 1000°C
- Up to 8 programmable ramp rates
- 25°C to 107°C, from 5 to 30°C/min
- 104°C to 1100°C, from 5 to 50°C/min
- Rotating carousel : holds up to 19 samples (metallic carousel)
- Metallic crucibles and covers
- Auto crucible-covers: automatic cover installation and removal (no oxidation)
- Pneumatic carousel mechanism
- Integrated precision balance with 0.1 mg precision readings
- Selectable atmospheres: N2/O2/Air
- Compressed air assisted cooling for faster cool down
- PC control: Windows® software
- Powerful ODBC data base
- Network for multiple TGA units (server option)
|
|
Advantages |
|
- The dual carousel configuration allows automatic installation and removal of crucible-covers inside the furnace.
- This prevents sample oxidizing and provides better volatile matter accuracy in addition to automation, (it works unattended).
- No danger to operator opening furnace at 600 ºC. No danger of covers falling covers furnace.
- Reduced analysis time in coal as the TGA does not go down from
900-950ºC to 600ºC to remove covers.
- Carousel, balance pedestal, and crucibles are made from a special alloy metal that all have a (2) year warranty.
- Temperature regulation is ± 1°C.
- Extremely robust mechanical design.
- Exceptional and un-surpassed reliability.
- Standard balance.
- Well known Eurotherm temperature controller.
- The modular chassis architecture reduces weight and makes it easier to access for service.
- Control by PC allows flexibility, ease of use and future modifications to the software will always be up to date.
|
|
Model Number |
Capacity |
|
|
TGA-2000 A |
19 Samples |
|
|
TGA-2000 AD |
38 Samples (dual furnace configuration) |
| |