Chlorine Controllers

 
 

Choosing the right controller needs to be decided based on the following criteria:

  • Parameters to be measured: pH, mV, T°C, Conductivity
  • Range
  • Control desired: ON / OFF or proportional
  • Set points: 1 or 2 for pH, 1 for other types of controllers
  • Is an alarm needed
  • Output signal: analog 4 to 20 mA to record or digital RS-232




PCA 300 and PCA 301 Chlorine Controllers


click on photo
to enlarge

 

Technical Data

 
  PCA 300U PCA 301U
Range 0-5 mg/L Free Chlorine

0-5 mg/L Total residual Chlorine

Precision ± 5% of the reading or  ± 0.05 mg/L (whichever is greater)
Resolution 0.01 mg/L
Minimum level of detection 0.05 mg/L
Repeatability ± 0.05 mg/L
Reagents Indicator HI 70450 (included)

Buffer HI 70451
(included)


DPD Compound HI 70452 (included)

Reagents will last for 50 days with a sampling rate of 10 min

Indicator HI 70460 (included)

Buffer HI 70461
(included)


DPD Compound HI 70452 (included)

Reagents will last for 50 days with a sampling rate of 10 min

Response Time

Depends on the sampling time.
A typical time is 5 minutes between 2 consecutive samples.

Sampling Rate Adjustable from 5 to 104 minutes
Sample Inlet Pressure From 1 psi (0.07 bar) to 20 psi (1.4 bar) with no external pressure regulator.  Pressure regulator is available for pressure exceeding 20 psi (1.4 bar).
Sample Flow Range Recommended flow rate: 300 mL/min.
Minimum allowed: 100 mL/min,  Maximum allowed: 500 mL/min
Sample Temperature Range 5 to 40°C
Interferences Oxidizing Agents: Iodine, Bromine, Ozone, Chlorine Dioxide, Permanganate, Hexavalent Chromium.
Hardness must not exceed 100 mg/L as CaCO3. Alkalinity must not exceed 400 mg/L for Free or 700 mg/L for Total Chlorine.
Operating Temperature 0 to 40°C
Recorder Output

Selectable: 0-10 mV, 0-100 mV, 0-1 V or 4-20 mA.
Output span can be set anywhere in the 0-5 mg/L range.

Proportional Relay/Alarms

1 Set Point:  Low or High, one system error alarm.
Each equipped with a  SPDT relay with contacts
(resistive load 5A at 250VAC or 5A at 30VDC),
(inductive load 2A at 250VAC or 3A at 30VDC)

Power Supply 20 VA at 115VAC/230VAC; 50/60 Hz.
Frequency must be specified at order.
Inlet Connection 3 mm ID tubing
Drain Connection 10 mm tubing
Case NEMA-4X molded fiberglass polyester
with transparent GE Lexan window
Dimensions / Weight

318 x 267 x 159 mm /
5 Kg without reagents


 